德国WEISS WK3-180/70气候模拟试验箱 技术规格书: 1.2 Technical Data WK3-180/70 1.2.1 Test Chamber Test chamber material stainless steel, 1.4301, surface III D polished Material of test space floor 1.4404, matt surface Test area dimensions height approx. 750 mm width approx. 580 mm width between shelf supports approx. 540 mm depth approx. 450 mm Volume of test chamber approx. 190 l Max. loading 130 kg Max. loading of test chamber floor 50 kg distributed load Max. load per shelf 30 kg distributed load Max. load of shelf supports 80 kg 1.2.2 Exterior Housing Material double-coated galvanised steel sheet Dimensions of housing height approx. 1805 mm width approx. 780 mm depth approx. 1385 mm Overall dimensions 1) height approx. 1805 mm width approx. 875 mm depth approx. 1545 mm Finish two-coloured, solvent-free, powder coated corpus grey-white (RAL 9002) front violet-blue (RAL 5000) Door hinged on the left, lockable, optimum contact pressure Feet adjustable, vibration absorbing 1 The required clearances can be reduced by dismounting components 1.3 Installation Drawing ? Stainless steel port included in basic design R1: approx. 125 mm ? L1: approx. 50 mm ? R21 … further installation position right L21 … further installation position left D11 installation position in ceiling 1. Main switch panel 2. 8” colour touch panel 3. CONTROLPAD 4. Condensation drain 5. Power supply cable length approx. 3,5 m 6. Door without window 8. Cooling water supply1 9. Cooling water drain1 10. Water supply (demineralised, H2O) 13. Slotted port1 14. Compressed air connection 1 # Width between shelf supports 1 Option 1.4 Equipment Shelf 1 pc., stainless steel Stainless steel ports 1 pc. approx. 125 mm ?, right side wall 1 pc. approx. 50 mm ?, left side wall each port incl. 1 silicone plug closed 1 slotted plug made of foamed silicone Condensation protection condensation cooler for prevention of condensation formation on specimen 8” TFT colour touch display fixed to front bracket CONTROLPAD integrated in door, for display of actual values and modification of operating state Control and programming digital measuring and control system S!MPAC* Interfaces ? 1 Ethernet interface (100/10 Megabit) for integration into network ? 1 USB interface for direct documentation of measured data via USB stick (USB stick not included in scope of supply) or as a printer interface Safety temperature limiter factory-set Specimen protection high/low temperature fuse according to EN 60 519-2, with separate sensor, thermal safety class 2 High temperature protection of humidification device high temperature controller Specimen switch-off potential-free contact Software temperature limiter tmin/tmax Digital I/O ? 4 digital outputs for control of customer-provided devices by means of potential-free contacts, load max. 24 V-DC, 0,5 A ? 4 digital inputs for feedback of customer provided devices, load max. 24 V-DC, approx. 30 mA